Battery Operated, Digital Vacuum Instrumentation measuring in mTorr, mBar and kPa
Features
Each Kurt Lesker vacuum gauge includes:
- A vacuum gauge controller
- A thermocouple vacuum gauge tube (vacuum sensor)
- A cable to connect the vacuum gauge controller to the thermocouple vacuum gauge tube
- NIST traceable calibration certificates may be purchased with initial order
- Gauges are calibrated for direct readout of nitrogen or air. Do not attempt to use with other gases unless accurate conversion data for N2 to the other gas is properly used.
Specifications Table
Model | 510TC |
---|---|
Technology |
Thermocouple
|
Measurements |
Torr Min: 1.000 x 10-3 Max: 1.33 mbar Min: 1.330 x 10-3 Max: 1.33 |
Temperature |
Operating 38.0 ºC 100 ºF Storage 38.0 ºC 100 ºF Bakeout 100 ºC 212 ºF |
Materials Exposed to Vacuum |
Glass Nickel Stainless Steel |
Typical Accuracy |
1 to 9 millitorr +/- 1 millitorr 10 to 499 millitorr +/- 10% of reading 500 to 1999 millitorr +/- 15% of reading |
Accuracy Range |
11 to 20%
|
Display Type |
Digital
|
Recommended Mounting Orientation |
Vertical
|
Input Power |
Battery
|
Compliance and Standards |
UL
|
Integrated Electronics |
Yes
|
Compatible Gauge Tubes |
DV-6M
KJL-6000SS |