Suggested Full Scale | 0.02 | 0.05 | 0.1 | 0.25 | 0.5 | 1 | 2 | 5 | 10 | 20 | 50 | 100 | 200 | 500 | 1000 |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Unit | Torr | Torr | Torr | Torr | Torr | Torr | Torr | Torr | Torr | Torr | Torr | Torr | Torr | Torr | Torr |
Total Measurement Range | 2.00E-02 to 2.00E-06 | 5.00E-02 to 5.00E-06 | 1.00E-01 to 1.00E-05 | 2.50E-01 to 2.50E-05 | 5.00E-01 to 5.00E-05 | 1.00E+00 to 1.00E-04 | 2.00E+00 to 2.00E-04 | 5.00E+00 to 5.00E-04 | 1.00E+01 to 1.00E-03 | 2.00E+01 to 2.00E-03 | 5.00E+01 to 5.00E-03 | 1.00E+02 to 1.00E-02 | 2.00E+02 to 2.00E-02 | 5.00E+02 to 5.00E-02 | 1.00E+3 to 1.00E-01 |
Minimum Suggested Measurement Pressure | 1.00E-05 | 2.50E-05 | 5.00E-05 | 1.25E-04 | 2.50E-04 | 5.00E-04 | 1.00E-03 | 2.50E-03 | 5.00E-03 | 1.00E-02 | 2.50E-02 | 5.00E-02 | 1.00E-01 | 2.50E-01 | 5.00E-01 |
Minimum Suggested Control Pressure | 1.00E-04 | 2.50E-04 | 5.00E-04 | 1.25E-03 | 2.50E-03 | 5.00E-03 | 1.00E-02 | 2.50E-02 | 5.00E-02 | 1.00E-01 | 2.50E-01 | 5.00E-01 | 1.00E+00 | 2.50E+00 | 5.00E+00 |
X CLOSE
Notes:
- It is suggested to not go below a 50 mV output for control.
- The following equation is an example for a 0-10 VDC output capacitance manometer. Please refer to the manual of your capacitance manometer to confirm the equation.
- Equation to convert output to pressure: p = (Uout / 10 V) x p (F.S.)
- ie. 0.05 V Uout for a 1000 F.S. unit : (0.05 V / 10 V) x 1000 = 5 Torr